32
CI/AZ40-EN Rev. A
Operating mode
In operating mode, process values from the connected sensor
are displayed on
Operator
Pages
– five
Operator
Pages
are
described in Sections (
Operator Page 1
) and (
Operator
Page 2
).
Operator Page 1
Operator
Page
1
displays 4 values simultaneously (Oxygen,
Combustibles, Inlet / Outlet / Outlet Temp. and Efficiency).
Operator Pages 2 to 5
Operator Pages 2
to
5
(Fig. 27.) each display a single value. Each
value (Oxygen, Combustibles, Inlet / Outlet / Outlet Temp. and
Efficiency) is associated with a template in the
Configuration
level
/
Display
/
Operator Templates
– see Operating instruction
OI/AZ40-EN. Minimum and maximum values are configurable in
the
Sensor Setup
level – see page 34.
*The highest priority diagnostic or alarm is displayed.
Other active diagnostic / alarm states can be viewed
on the
Diagnostics View
– see page 33.
Fig. 26 Operator page 1 (4 process values displayed)
20.9
0.06
97
0.0
CAL
2015-04-09
14:33:12
°C
%
%
%
Instrument tag
(user-assignable)
Date and time
Soft key icon
Autoscroll
icon –
scrolls
Operator
Pages
sequentially
when enabled
Oxygen
Combustibles
Oxy. High Alarm
AZ40
Outlet Temp.
Efficiency
NAMUR (NE107) Status Icons
Failure
High Process Alarm
Maintenance
Low Process Alarm
Out of Spec
High Latch Alarm
Check Function
Low Latch Alarm
Diagnostic message or alarm tag and diagnostic message*
Process output description
(color-coded tag)
Process value
Operator Menu
icon
Title bar icon
Diagnostic message
IMPORTANT (NOTE)
If the measured value is above the specified range, the
(color-coded) bargraph flashes to indicate an excess
value for the displayed process.
Fig. 27 Operator pages 2 to 5
%
CAL
20.9
2015-04-15
15:31:08
Oxygen
AZ40
Instrument tag
(user-assignable)
Value for displayed
process output
Unit of
measure
Operator Menu
icon
Color-coded bargraph
Oxygen = red
Combustibles = green
Inlet / Outlet / Outlet Temp. = blue
Efficiency = violet
Color-coded output tag
Oxygen = red
Combustibles = green
Outlet Temp. = blue
Efficiency = violet
Diagnostic icon /
diagnostic message
Oxy. High Alarm
CAL
(calibration)
shortcut prompt
Autoscroll
icon
Date and
time