21
To stop during A-BPM
When the START/STOP switch is pressed during blood
pressure measurement, the air is expelled immediately and the
current measurement is stopped. However, A-BPM is continued.
The next blood pressure measurement is performed in
accordance with A-BPM settings.
To set the program for A-BPM.
Step 1. Set the AUTO switch to "
OFF
".
Step 2. If the display is hidden,
press the START/STOP or EVENT switch to return to the
waiting mode display.
Step 3. While pressing and holding the START/STOP switch,
press and hold the EVENT switch until Sleep is displayed
on the OLED.
Step 4. The operation switches are as follows:
Refer to "
8.3.1. A-BPM Items and Parameters
"
EVENT switch ................. Change the current parameter.
START/STOP switch ...... Decision, next item, end of settings.
To measure blood pressure during A-BPM immediately.
(Manual blood pressure measurement of A-BPM)
Step 1. If the OLED is hidden, press the
START/STOP or EVENT switch to return to the A-BPM
waiting mode display. The A-BPM
waiting mode
is a state
when blood pressure is not measured during the
interval
time
.
Step 2. Press the START/STOP switch during the A-BPM waiting
mode.
Содержание TM-2441
Страница 20: ......
Страница 36: ...16 4 2 Dimensions 66 24 5 95 Unit mm...
Страница 89: ...69 Air hose Air hose Cuff Belt Carry holder Air plug Air socket Insert Carry holder Belt Tape Clip...
Страница 119: ......