43
7.3.
Inspection for Use
Inspect the recorder to maintain its performance, safety and
effectiveness before use.
Confirm the following checklist before/after installing the
batteries.
If a problem is found, stop using the recorder and attach a
"
Malfunction
" or "
Do
not use
" message. Contact your local
dealer to repair it.
7.3.1.
Battery Pre-installation Checklists
No.
Item
Description
1 Exterior
No damage and deformation from dropping.
No damage and looseness of switches, etc.
2 Battery
Check that the batteries are not flat. Replace with
two new batteries before the patient uses it.
3 Cuff
Check that the cuff has not frayed. If the cuff is
frayed, it may burst due to internal pressure.
4
Cuff
connection
Check that there are no kinks and folding in the air
hose.
Check that the air socket and connector is
connected firmly.
5 Attachments
Check that there is no damage to accessories.
(Carry holder, belt, etc.)
Caution
Содержание TM-2441
Страница 20: ......
Страница 36: ...16 4 2 Dimensions 66 24 5 95 Unit mm...
Страница 89: ...69 Air hose Air hose Cuff Belt Carry holder Air plug Air socket Insert Carry holder Belt Tape Clip...
Страница 119: ......