Automated night blood pressure display
ANBP
The ANBP can specify up to six preset
start times
for the
"
1 scope
". Refer to "
6.2.1. S-BPM Programs
" for the "
1 scope
".
When the last
start time
is the same as first
start time
, the
settings for
start time
finish. Next, specify the
measurement count
and
interval time
for the "
1 scope
".
First
start time
2nd
start time
3rd
start time
Measurement count
Interval time
Example Settings and simplified input
The measurement happens at 22:00 and 4:00.
The
start time
of section 1 =
4:00
The
start time
of section 2 = 22:00
The
start time
of section 3 = 4:00 ................. The same as section 1
The
measurement count
=
2 times
The
interval time
=
0:01 minute