89
Code
Meaning
Cause and treatment
Excessive body
motion.
Relax and keep still during
measurement.
Out of range,
30 PUL 200 If these errors occur multiple times, try
another blood pressure measurement.
#1
PP = SYS - DIA
SYS: Systolic blood pressure
DIA: Diastolic blood pressure
PP: Pulse pressure
Out of range,
30 DIA 160
Out of range,
60 SYS 280
Out of range,
10 PP 150
#1
Measurement is
above 180 seconds.
If the inflation speed or exhaust speed is
slow, an inspection is necessary.
Exhaust is above
90 seconds.
The exhaust speed may be slow,
an inspection is necessary.
Heartbeat cannot be
detected.
Heartbeat cannot be detected
because of body motion, etc.
Measure the blood pressure while
relaxed and do not move.
The settings of the
interval time are
incorrect.
If the interval time is set to 120 minutes,
the difference between last
start time
and next
start time
cannot divide into
two hours perfectly.
Zero pressure error
for safety circuit.
Displays at the measurement start
time.
Release the air remaining in the cuff
completely.