Blood Pressure Measurement Operations
8.7.1.
A-BPM Operations
When A-BPM is started, the blood pressure is measured in
accordance with the preset parameters.
Note
Set the built-in clock and initial pressurization value before
measurement, as A-BPM uses them. Refer to "
8.2.2. The Clock
and the Monitor Function of Measurement
" and "
8.3. A-BPM
Preset Programs
".
When the recorder is removed, set the AUTO switch to "
OFF
".
If the recorder is removed during A-BPM, the inflation of the
cuff starts at the next
start time
and the cuff may burst.
When A-BPM is resumed, set the AUTO switch to "
ON
".
The
mark is displayed while A-BPM is used.
Manual blood pressure measurement can be performed during
A-BPM waiting mode.
The measurement result for the manual blood pressure
measurement can be stored in the memory.
When A-BPM is stopped, the error code E07 is displayed on
the OLED and stored in the memory.
To start A-BPM
Step 1. Set the AUTO switch to "
ON
".
Step 2. The
mark is shown on the OLED and LCD. A-BPM is
started.