72
8.7.2.
S-BPM Operations
Note
Set the built-in clock and initial pressurization value before
measurement because they are used for S-BPM.
Refer to "
8.2.2. The Clock and the Monitor Function of
Measurement
" and "
8.4. S-BPM Programs
".
Remove the batteries when the patient removes the recorder and cuff
when ANBP or ASBP is used (even during the waiting mode). If the
batteries are left in the battery compartment, the cuff may break when
the recorder inflates the cuff at the next "
1 scope
". If the patient resumes
measurement, insert the batteries and press the / switch.
Manual blood pressure measurement can be performed during
S-BPM waiting mode.
The measurement result for the manual blood pressure
measurement can be stored in the memory.
When S-BPM is stopped, the error code E07 is displayed on the
OLED and stored in the memory.
To start S-BPM
Step 1. Set the AUTO switch to "
OFF
".
Step 2. Operations are as follows:
S-BPM Programs
Operations
Office blood pressure
OBP Press the / switch to
start the preset program
during waiting mode.
Automated office blood pressure AOBP
Home blood pressure
HBP
Automated night blood pressure ANBP Preset program enters
standby until the "
start time
"
or "
start time of alarm
".
Automated self-blood pressure
ASBP