53
#1
:
When the sleep mode is set to "
ON
", the
start time
and
operation
time
of the automated measurement and the
interval time
of the
sleep mode can be used. The
interval time
for these sections
(1 to 6) cannot be used.
#2
:
When the sleep mode is set to "
OFF
", the
interval time
for the sleep
mode is not displayed.
#3
:
If the
start time
is specified and the
operation time
is set to
"
OFF
", when the AUTO switch is set "
ON
", the
automated
measurement
starts at the preset
start time
and continues
until the AUTO switch is set to "
OFF
". If the AUTO
switch
is set to "
ON
" again, the
automated measurement
starts at
the preset
start time
.
Note
When the
operation time
is specified, even if the AUTO
switch is operated during the
automated measurement
, the
automated measurement
continues for the
operation time
from the time that the AUTO switch is initially set to "
ON
".
#4
:
If the
start time
is set to "
OFF
" and the
operation time
is
specified, when the AUTO switch is set to "
ON
", the
automated measurement
performs the first blood pressure
measurement and continues for the
operation time
.
If the AUTO switch is set to "
OFF
" during the
automated
measurement
, it stops.
If the AUTO switch is set to "
ON
" again, the
automated
measurement
is performed for the
operation time
.
Содержание TM-2441