56
Example 2 Automatic measurement
The
start time
for the automated measurement = 7:00,
The
operation time
for the automated measurement = OFF.
When the AUTO switch is set to "
ON
", A-BPM is started at 7:00.
A-BPM is continued according to the
start time
and
interval
of each
section until the AUTO switch is set to "
OFF
".
#
Even if the AUTO switch is set to "
OFF
" and to "
ON
" again
during the
automated measurement
, the
automated
measurement
continues.
Example 3 Automatic measurement
The
start time
for the automated measurement = OFF,
The
operation time
for the automated measurement = 26 hours.
When the AUTO switch is set to "
ON
", A-BPM continues
according to the
start time
and
interval
for each section for
26 hours.
AUTO switch to "
ON
"
Blood pressure measurement
Time
Section 1
After 26 hours
Section 2
Not measured
AUTO switch to "
ON
"
Blood pressure measurement Not measured
Time
Section 1
AUTO switch to "
OFF
"
Section 2
7:00