23
5.3.2.
S-BPM Operations
To start S-BPM.
Step 1. Select the S-BPM program and store its parameters.
Step 2. Set the AUTO switch to "
OFF
".
Step 3. Operations are as follows:
S-BPM Programs
Operations
Office blood pressure
OBP Press the / switch to
start the preset program
during waiting mode.
Automated office blood pressure AOBP
Home blood pressure
HBP
Automated night blood pressure ANBP Preset program enters
standby until the "
start time
"
or "
start time of alarm
".
Automated self-blood pressure
ASBP
To stop S-BPM.
Operations are as follows:
S-BPM Programs
Operations
Office blood pressure
OBP
Press the / switch to stop blood
pressure measurement.
Automated office
blood pressure AOBP
Home blood pressure
HBP
Automated night
blood pressure ANBP
Press the / switch to stop blood pressure
measurement. At the next start time, the blood
pressure is measured or the buzzer sounds.
#1
If you need to stop the recorder completely,
remove the batteries from the recorder or
switch to OBP, AOBP, or HBP.
Automated self-
blood pressure ASBP
#1
:
Refer to "
6.2.1. S-BPM Programs
".