28
The
mark is displayed on the LCD while A-BPM is used.
Blood pressure is measured automatically at A-BPM start time.
An initial pressurization value of A-BPM is AUTO, so that
a proper pressurization value is selected automatically.
If the first pressurization is insufficient, re-pressurizations are
performed automatically up to two times.
When you delete data from the memory or move the AUTO
switch to "
OFF
", the pressurization value is reset to the initial
pressurization value.
When a measurement error occurs and the waiting time until
the next start time is longer than 8 minutes, blood pressure is
measured once after 120 seconds. The measurement result is
stored in the memory.
If you want to suspend A-BPM, release the LOCK switch and
move the AUTO switch to "
OFF
".