5.Trigger Setup - 5.3.Pre-Trigger
1WMPD4004444
47
Description of trigger source (5)
Trigger filter
The trigger filter function ensures that a trigger is detected when the trigger conditions are met for a
specified period of time, in order to prevent erroneous trigger detection due to noise or chattering in the
signal near the trigger level.
Trigger detection point
Trigger level
CHn input signal
T Trigger filter
5.3.
Pre-Trigger
When performing memory recording, a pre-trigger can be set to adjust the recording length before and
after the trigger detection point in the memory block.
Trigger detection
Trigger level
End measurement
Start measurement
Input signal
Pre-trigger
(post-trigger)
30%
(70%)
Memory block
5.3.1. Pre-Trigger Setup
Tap the
【
SETUP
】
key on the side menu to display the setup menu.
Tap the
【
Recording
】
tab in the recording setup to display the recording setup screen.
The recording device settings are displayed below the recording setup.
Tap
【
Pre-Trigger
】
on the right of memory recording to set the pre-trigger.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...