11.Maintenance -11.10.Troubleshooting and Inspection
1WMPD4004444
141
11.10. Troubleshooting and Inspection
If this product does not operate normally after performing the indicated countermeasure or a repair is
required, please contact our sales representative.
Symptom
Possible cause
Countermeasure
The power does not
turn on.
Nothing is displayed
on the screen.
The power cord is not connected to
the connector properly.
Connect the power cord properly and turn
on the power switch
The fuse has blown.
The fuse for this product cannot be replaced
by the customer because it is located inside
the main unit. Please contact our sales
representative if the fuse may be blown.
The screen is set to turn off
automatically.
Press any key to turn on the screen.
The touch panel or
keys do not respond.
The product is recording.
The start LED is on.
Perform the operation again after pressing
stop on the operation panel to stop
measurement.
The key lock is enabled.
Turn off the side touch panel lock.
Printer recording is not
performed.
There is no recording paper.
Load recording paper.
The printer cover is open.
Close the printer cover.
The thermal head is at an abnormally
high temperature.
Use the product in a location at a
temperature between 0 and 40°C. Do not
continuously print solid black areas.
Real-time waveform printing is
disabled in the recording settings.
Enable real-time waveform printing in the
recording settings and start recording.
SSD recording is not
performed.
There is no free space remaining on
the SSD.
Delete unnecessary recorded data.
Memory recording
cannot be replayed.
Memory data is not saved because
the trigger is not enabled.
Cause a manual trigger using the TRIG key
on the operation panel.
Recording does not
start when the start
key is pressed.
The recording mode is set to trigger
start or time start.
Disable the START trigger.
External synchronization recording is
enabled.
Press the start key after inputting the signal,
as recording cannot start unless a pulse
signal is input to the remote terminal.
Recording paper has not been
loaded.
Load recording paper.
The key lock is enabled.
Disable the side key lock.
Data cannot be saved
to the specified media.
The media has not been formatted. Format the media.
There is insufficient free space on the
media.
Delete unnecessary files or use new media.
The media is set to read-only.
Disable the read-only setting of the media.
The media is not
recognized.
The format of the media is invalid.
Format the media.
The media is damaged.
Use other media.
The communication
interface cannot be
used to configure
settings and control
operation.
The communication parameter
settings do not match.
Match the address and communication
parameters.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...