9.5.4. Reference Materials
CAUTION
Do not connect something other than a sensor (thermocouple (TC) or platinum resistance temperature
detector) to an input terminal.
(Do not input voltage or current. Doing so may lead to failure.)
Ensure that the voltage between each input and the chassis (GND) and between each channel does
not exceed 300 V AC or DC. (Damage may be caused if the voltage is exceeded.)
When measuring a location with electric potential using a non-insulated thermocouple (TC) or platinum
resistance temperature detector, never touch the metallic parts of the input wire, as electric potential
occurs in the cable itself.
1 Front panel
2 Thermocouple (TC) sensor connection method
2.1 Connection terminal
After screwing the thermocouple (TC) into the temperature sensor connector, insert it into the connector
of the front panel.
The terminal names of the input connectors are indicated on the left side of the panel as + , - ,
and S .
+ /A terminal
Connect the + side wire of the thermocouple (TC).
- /B terminal
Connect the - side wire of the thermocouple (TC).
S /b terminal
Terminal for shielding. Connect the sheath (shield) wire of the non-contact type
sheathed thermocouple (TC). Leave the S terminal unconnected for the grounded
type.
NOTE
When using the non-contact type sensor, the S terminal must be insulated from the + terminal and -
terminal.
If they are connected, correct measurement will not be able to be performed and failure may occur.
The thermocouple (TC) has a + and - side. Connect them correctly.
(Correct measurement will not be able to be performed if the + and - sides are switched.)
Make sure the resistance value of the thermocouple (TC) is 1 kΩ or less.
Use a compatible extension wire for thermocouple (TC) extension.
Connect a single thermocouple (TC) to a single channel.
Wire the thermocouple (TC) cable away from the power line and sources of noise.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...