5.Trigger Setup - 5.4.Start Trigger
48
1WMPD4004444
Standby state waiting
for START trigger to
start recording
5.4.
Start Trigger
The start trigger function starts recording when the trigger conditions are established for the channel
specified in the trigger source. Press the START key on the operation panel to put the product in the
standby state.
Printer recording, SSD recording, and memory recording start when the start trigger is detected.
Press START key
Start trigger occurs
Recording ends
Recording time
During record
Printer recording
SSD recording
Memory recording
Trigger occurred Trigger occurred Trigger occurred
5.4.1. Start Trigger Setup
Step 1. Tap the CH key on the side menu to display the channel setup sub menu.
Step 2. Tap the Trigger tab on the bottom right of the sub menu to display the trigger setup screen.
Step 3. Tap the trigger menu selection on the top of the trigger setup screen (1) to display [Start trigger].
(1) Trigger menu selection:
Switches the menu.
(2) Channel: Selects the start trigger source channel.
(3) Detection: Selects UP, DOWN, INTO WIN, or OUT WIN for the
polarity of the trigger signal.
(4) Threshold: Sets the trigger level (threshold).
(5) Filter:
Sets the filter time for noise removal.
The setup and operations of (1) to (5) are the same as the memory
trigger on page 45.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...