12.Specifications - 12.1.General Specifications
144
1WMPD4004444
*
1
The Pollution Degree indicates the level of pollution that can exist in the ambient environment.
Pollution degree 1:
No pollution or only dry, non-conductive pollution occurs. The pollution has no
influence.
Pollution degree 2:
Only non-conductive pollution occurs except that occasionally a temporary
conductivity caused by condensation is to be expected.
Pollution degree 3:
Conductive pollution occurs, or dry, non-conductive pollution occurs which
becomes conductive due to condensation which is to be expected.
Pollution degree 4:
The pollution generates persistent conductivity caused by conductive dust or by
rain or snow.
*
2
The Overvoltage Category (Installation Category) indicates how much overvoltage (impulse voltage)
from an AC power supply an electrical device can withstand. Overvoltage Category II (CAT II) is suitable
for devices powered by wire from the switchboard of a building.
*
3
The Measurement Category categorizes a testing or measurement circuit according to the type of main
power circuit intended to be connected for testing or measurement, and differs according to the modules
installed to this product. Use the product within the Measurement Category that meets the module
specifications.
CAT II:
Applies to testing and measurement circuits directly connected to the point of use
(power outlet or similar location) of a low voltage main power supply system.
CAT III:
Applies to testing and measurement circuits connected to the switchboard of the low
voltage main power supply system in a building.
CAT IV:
Applies to testing and measurement circuits connected to the source of the low
voltage main power supply system in a building.
No category (O): Applies to circuits not directly connected to a main power supply.
Switchboard
Devices directly powered by switchboard
CAT IV
Power outlet
CAT III
Devices powered
by outlet
CAT II
Devices not
directly connected
to main power
O
O
: Other circuits not directly connected to main power
CAT II : Measurement Category II
CAT III : Measurement Category III
CAT IV : Measurement Category IV
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...