8.Setup Details - 8.1.Recording Setup
1WMPD4004444
79
Disp. pos.: Displays the display position set in the channel settings. Tap to change the display position.
Disp. range: Displays the display range set in the channel settings. Tap to change the display range.
Disp. max: Displays the display maximum set in the channel settings. Tap to change the display
maximum.
Disp. min: Displays the display minimum set in the channel settings. Tap to change the display minimum.
Conversion (Physical Quantity Conversion)
The items in the list are indicated below.
Batch:
Enables the batch configuration of the selected channels.
You can tap Select all on the top right to select all the channels.
Conversion method: Select [Gain], [2-pt.], or [None] as the conversion method.
Gain:
Applies gain and offset to the input voltage and performs conversion with the primary
function y = ax + b.
Conversion1 is gain, Conversion2 is offset
2-pt.:
Converts two input voltages to two target physical quantities.
Conversion1 is the first physical conversion value, and Conversion2 is the second
physical conversion value
For example, when converting the voltage of a signal input at 4-
20 mA with 1 kΩ
shunt resistance, the setting for resisting the input voltage value at ±1 V is
Conversion1 = 20 → +1, Conversion2 = 4 →
-1.
None:
Do not perform physical quantity conversion.
Relationship between 2-pt gain and gain compensation
The relationship between the gain specified by the 2
points a1 (x1, y1) and a2 (x2, y2) and gain a and offset b
of y = ax + b is
gain a = (y2 - y1)/(x2 - x1)
offset b = y1 - ax1 = y1 - x1(y2 - y1)/(x2 - x1)
Physical quantity
y = ax+b
y2
a2
y1
a1
b
a
x1
x2Input value
Unit:
Sets the unit for the output of the physical quantity conversion. Tap to display the unit table and
select the target unit. If the target unit is not in the table, tap the Unit list key and change the
unit table.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...