4.Configuring Measurement - 4.2.Setup the Input Channel
38
1WMPD4004444
4.2.
Setup the Input Channel
4.2.1. Channel setup sub menu (for RA30-101)
(1) Slot number, input module type
(2) Change slot:
You can change the display slot by swiping
this sub menu left or right or tapping the left
< and right > key.
(3) Select channel:
Select the channel in the slot.
(4) Meas. ON/OFF
ON:
Measure and record the input signal.
(5) Color:
Change the display color of the
waveform monitor.
(6) Coupling: Switch the input signal coupling in the
order DC → GND → AC.
(7) Meas. range:
Change the measurement range of the
input channel.
When this key is tapped, the rotary knob
is enabled (the LED lights up) and the
range can be changed by turning the
knob.
(1)
(2)
(2)
(3)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
(16)
(8) L.P.F.:
Change the low-pass filter of the input channel. When this key is tapped, the rotary knob is
enabled (the LED lights up) and the filter can be changed by turning the knob.
(9) A.A.F.:
Turns the anti-aliasing filter of the input channel on or off.
(10) Disp. pos.: Specify the position of the waveform monitor to display the specified range of the
waveform display area. Specified as a percentage indicating the center position of the
display range from the bottom of the monitor when the full range of the monitor is 100%.
(11) Disp. range: Specifies the display width in the amplitude direction of the waveform display area on
the waveform monitor.
Specified as the percentage of the display width when the full range of the monitor is 100%.
Example) When 50% is set, the waveform display is 10 div of the total width of 20 div.
(12) Disp. min: Set (by tapping the key and turning the knob) the display lower limit value (scale value)
of the bottom of the display range.
(13) Disp. max: Set (by tapping the key and turning the knob) the display upper limit value (scale value)
of the top of the display range.
(14) Zero adjust: Cancels the input offset of the input channel. Execute zero adjust to perform more
accurate measurement.
(15) Sheet:
Set the monitor display/printer print sheet of the set channel.
(16) The available range and current measurement value monitor.
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...