9.Using Optional Modules - 9.6.Remote Control Module (RA30-112)
124
1WMPD4004444
9.6.6. Reference Materials
1
Front panel
2 Pin layout of remote input connector
Pin
number
Sig
name
Pin
number
Sig
name
Function
A1
(NC)
B1
IGND
-
A2
(NC)
B2
IGND
-
A3
START/
STOP
B3
IGND
Measurement signal/stop signal
Performs measurement for the period where the signal is at
the L level.
A4
MARK IN
B4
IGND
Mark
Records a mark in the recorded data when the signal
becomes L during recording.
A5
FEED IN
B5
IGND
Feed (idle feeding of recording paper)
Performs idle feeding of printer recording paper for the
period where the signal is at the L level.
A6
PRINT IN
B6
IGND
Performs printer recording for the period where the signal is
at the L level while stopped. This signal is disabled during
recording.
A7
(NC)
B7
IGND -
A8
EXT
SMPL IN
B8
IGND
(SMPL)
External synchronization sampling
The data sampling and printer recording of this product is
performed with this sampling clock.
A9
TRIG IN
B9
IGND
(TRIG)
Trigger
The falling of this signal operates as trigger detection.
The memory trigger and START trigger can be switched.
A10 ACCLK IN
B10
IGND
(ACCLK)
OSC signal for AC strain module
This signal is distributed to the installed AC strain module
as the excitation signal for the AC strain module.
Connector for input signal: DF02R020NA3 (Japan Aviation Electronics Industry)
The A series and B series are complementary. IGND (B series) is the input common for A series signals.
IGND is shared with input/output terminals inside the module.
IGND is isolated from the main unit frame.
The plugs (manufacturer model numbers) corresponding to each input signal connector are indicated
below.
Compatible plug: DF02P020F22A1 (soldered type), DF02P020G28A1 (pressure connected type)
Содержание RA3100 Omniace
Страница 1: ...RA3100 Omniace Instruction Manual 1WMPD4004444...
Страница 162: ...13 Optional Parts 13 1 List of Cables 162 1WMPD4004444 12 4 Exterior 12 4 1 Main Unit Exterior 394 199 410 334...
Страница 163: ...12 Specifications 12 4 Exterior 1WMPD4004444 163 12 4 2 Two Channel Voltage Module RA30 101 Exterior...
Страница 164: ...13 Optional Parts 13 1 List of Cables 164 1WMPD4004444 12 4 3 Four Channel Voltage Module RA30 102 Exterior...
Страница 165: ...12 Specifications 12 4 Exterior 1WMPD4004444 165 12 4 4 Two Channel Voltage Module RA30 103 Exterior...
Страница 166: ...13 Optional Parts 13 1 List of Cables 166 1WMPD4004444 12 4 5 16 Channel Logic Module RA30 105 Exterior...
Страница 167: ...12 Specifications 12 4 Exterior 1WMPD4004444 167 12 4 6 Two Channel Temperature Module RA30 106 Exterior...
Страница 168: ...13 Optional Parts 13 1 List of Cables 168 1WMPD4004444 12 4 7 Remote Control Module RA30 112 Exterior...
Страница 173: ...13 Optional Parts 13 2 List of Probes Clamp Meters 1WMPD4004444 173 MEMO...
Страница 174: ...13 Optional Parts 13 4 List of Spare Parts 174 1WMPD4004444 MEMO...
Страница 175: ...Omniace RA3100 Instruction Manual 1WMPD4004444 1st Edition...