2.7 Risk of fire
000000-2096-101-DokS-GB-181218
19 / 116
2.7 Risk of fire
Do not use the laser in conjunction with flammable anesthetic
substances.
Keep potentially explosive materials away from the laser hazard
zone. Easily flammable materials may start fires.
A laser beam could ignite many explosive or flammable gases or
liquids, including some of the solutions used in the preparation of
surgical procedures.
Flammable drapes, surgical gowns, gauze and other ignitable
materials must be kept out of the beam path. We recommend
using non-flammable materials and instruments and wearing
gowns and clothing which do not easily catch fire.
A fire extinguisher should be kept in the vicinity of the system.
Check the integrity of all lenses and glass surfaces before each use.
Severe contamination may result in unwanted heat exposure.
Detached coatings may result in exposure to increased radiation.
2.8 Damage to fiber optics
Fiber optics may not be tightly bent, kinked or incompletely
secured. The system may otherwise be damaged and injury caused
to the patient or user.
2.9 Maintenance measures
Maintenance procedures (maintenance and repairs) which are not
specified in these instructions for use may only be carried out by
persons authorized by Carl Zeiss Meditec and solely according to
the service instructions issued by Carl Zeiss Meditec. For planning
and implementing these maintenance and care procedures please
contact ZEISS Service or your local dealer.
Summary of Contents for VISULAS green
Page 1: ...VISULAS green Documentation set...
Page 20: ...Empty page for your notes...
Page 42: ...Empty page for your notes...
Page 48: ...Empty page for your notes...
Page 92: ...Empty page for your notes...
Page 110: ...Empty page for your notes...
Page 112: ...Empty page for your notes...
Page 114: ...Empty page for your notes...