2 Safety notes
VISULAS green
2.5 Risk from incoherent optical radiation
000000-2096-101-DokS-GB-181218
15 / 116
• on easily inflammable materials,
• in explosion risk areas (e.g. combustible mixture of anesthetic,
cleaning or disinfecting agents with air, oxygen or nitrous
oxide).
u
Do not store or use this system in damp areas. Do not expose
the system to water splashes, dripping or sprayed water.
u
Ensure that no liquids can enter the system.
Decommissioning
u
If one of the following events should occur, disconnect the
cable from the power supply, label the system clearly as being
out of service and report the problem to the ZEISS service:
• Electric shocks
• Penetration of substances
• Frequently occurring error messages
• Faults that cannot be remedied based on the information
provided in these instructions for use
Symbols and labels
u
Observe the symbols and labels attached to the system.
Transport
u
Transport the system over long distances (e.g. move, return for
repair, etc.) only in original packaging or special return
packaging.
u
Contact your dealer or ZEISS Service.
2.5 Risk from incoherent optical radiation
Because prolonged intense light exposure can damage the retina,
the use of LSL green or LIO VISULAS green for ocular examination
should not be unnecessarily prolonged, and the brightness settings
should not exceed what is needed to provide clear visualization of
the target structures.
The halogen or LED light emitted from these devices is potentially
hazardous. The longer the duration of exposure, the greater the
risk of ocular damage. Exposure to light from these devices when
operated at maximum output will exceed the safety guideline after
8 min (filter position: white).
If you have any questions concerning permissible longer exposure
times with other filter positions, please contact ZEISS service.
When operating the devices, avoid looking directly into the light
emitting prism.
Take particular care, when examining infants and aphakic patients.
Never look into the sun through the binocular tube and eyepieces.
Summary of Contents for VISULAS green
Page 1: ...VISULAS green Documentation set...
Page 20: ...Empty page for your notes...
Page 42: ...Empty page for your notes...
Page 48: ...Empty page for your notes...
Page 92: ...Empty page for your notes...
Page 110: ...Empty page for your notes...
Page 112: ...Empty page for your notes...
Page 114: ...Empty page for your notes...