8 Cleaning and disinfection
Documentation set
VISULAS green
8.3 Disinfection
84 / 116
000000-2096-101-DokS-GB-181218
u
To clean the monitor, commercially available cleaning cloths for
computers and monitors should be used.
u
Use a commercial lens cleaning tissue (moist or dry) for
cleaning the optical surfaces on the patient's side.
u
Remove dust from optical surfaces by means of a fine brush.
u
Use the supplied dust cover to protect the laser slit lamp from
dust when not in use.
8.3 Disinfection
Instrument parts that routinely contact the patient, i.e. the
forehead rest and chinrest, should be disinfected before each
examination. These parts are not removable.
We recommend wipe disinfection, since the optical components
could be adversely affected by spraying of disinfectants. Please
note that only the forehead rest and chinrest should be disinfected,
not any of the optical components of the device.
Action
u
In case of coarse contamination (e.g. makeup or dirt stains)
remove visible stains on the forehead rest and chinrest with a
soft cloth, moistened with a mild detergent or 70 % isopropyl
alcohol solution.
u
Then use another soft cloth, moistened with 70 % isopropyl
alcohol solution and thoroughly wipe the surfaces of the
forehead rest and the chinrest.
u
Allow the surfaces to dry completely, before the patient comes
in contact with it.
Summary of Contents for VISULAS green
Page 1: ...VISULAS green Documentation set...
Page 20: ...Empty page for your notes...
Page 42: ...Empty page for your notes...
Page 48: ...Empty page for your notes...
Page 92: ...Empty page for your notes...
Page 110: ...Empty page for your notes...
Page 112: ...Empty page for your notes...
Page 114: ...Empty page for your notes...