11 Technical specifications
Documentation set
VISULAS green
11.6 Electromagnetic compatibility
102 / 116
000000-2096-101-DokS-GB-181218
Guidance and manufacturer’s declaration – electromagnetic immunity
The VISULAS green is intended for use in the electromagnetic environment specified below. The customer or
the user of the VISULAS green should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Conducted HF disturbances
pursuant to IEC 61000-4-6
including lines with a maximum
length of less than 3 m (Note b in
Table 8, IEC 60601-1-2:2014)
3 V
RMS
150 kHz to 80 MHz
3 V
6 V
RMS
ISM bands
a
6 V
Radiated HF IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Radiated HF disturbances by near
fields of wireless communication
according to IEC 61000‑4‑3
27 V/m
380 MHz to 390 MHz
27 V/m
28 V/m
430 MHz to 470 MHz,
800 Mhz to 960 MHz,
1.7 GHz to 1.99 GHz,
2.4 GHz to 2.57 GHz
28 V/m
9 V/m
704 MHz to 787 MHz
5.1 GHz to 5.8 GHz
9 V/m
Note 1: At 80 MHz and 800 MHz the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795
MHz; 13.553 MHz to 13,567; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Summary of Contents for VISULAS green
Page 1: ...VISULAS green Documentation set...
Page 20: ...Empty page for your notes...
Page 42: ...Empty page for your notes...
Page 48: ...Empty page for your notes...
Page 92: ...Empty page for your notes...
Page 110: ...Empty page for your notes...
Page 112: ...Empty page for your notes...
Page 114: ...Empty page for your notes...