Optional accessories
000000-2121-813-GA-GB-050416
48
Optional accessories
WARNING - RISK OF OPTICAL RADIATION
Use only accessories and spare parts approved by Carl Zeiss Meditec.
WARNING - RISK OF ELECTRIC SHOCK
If connecting external components to the interfaces of the device, the
operator must ensure to meet the safety requirements as per
IEC 60601-1:2005, chapter 16 (medical electrical systems)!
For connection to a LAN, a network isolator must be installed (cut-off
voltage at least 4 kV). To ensure electromagnetic immunity, only network
isolators supplied by Carl Zeiss Meditec may be used. Network isolators and
network cables must be installed in accordance with Fig. 18.
A USB isolator (cut-off voltage at least 4 kV) with USB cable must be used
for data exchanged via external USB devices with separate power supply.
The USB isolator may not be powered by a separate power supply.
CAUTION - RISK OF OPERATING ERRORS
Only software authorized by Carl Zeiss Meditec may be used in this
instrument.
The device may only be connected to private networks which are protected
from public networks (Internet) by firewalls conforming to the latest
technical standards!
The following may be purchased as optional accessories for CIRRUS photo:
•
Demonstration eye
•
USB flash drive
•
Printer
•
Network isolator and network cable
•
USB isolator and USB cable
A complete up-to-date list of accessories can be obtained from your dealer.
Summary of Contents for CIRRUS photo 600
Page 1: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 User manual...
Page 8: ...000000 2121 813 GA GB 050416...
Page 73: ......
Page 76: ......
Page 77: ...Remote maintenance tool Addendum to the documentation set...
Page 78: ...000000 2121 813 AddGA01 GB 210915...
Page 85: ......
Page 88: ......
Page 104: ...WINDOWS EMBEDDED STANDARD 7 GB 21 06 2012 16...
Page 105: ......