Technical data
000000-2121-813-GA-GB-050416
55
Imaging
Fundus camera image capture
Field angle
45° and 30°
Ametropia compensation
+35 D to -35 D continuous
Filter
Blue/IR cut filter (UV and IR protection)
Fluorescein and ICG angiography filter
Fundus autofluorescence filter
Max. image sequence
1 image every 2 s
Magnification
0.5x on image sensor (image diameter on sensor 7.4 mm),
independent of focusing state
Light source
Xenon flash bulb
HD-OCT imaging
Axial resolution
5 µm (in tissue)
Transversal resolution
15 µm (in tissue)
Scan rate
27,000 A-scans per seconds
A-scan depth
2.0 mm (in tissue); 1024 pixels
Radiation source
Superluminescence diode (SLD), 840 nm
Optical power
< 725 µW on the cornea
Ambient conditions for intended use
Temperature
+10
°
C to +35
°
C
Relative humidity
30 % to 90 %
Altitude
up to 3,000 m above sea level
Environmental conditions for storage and transport (unpacked)
Temperature
-10 °C to +55 °C
Relative humidity
10 % to 95 %
Ambient conditions for storage and transport in original packaging
Temperature
-40 °C to +70 °C
Relative humidity
10 % to 95 %
Summary of Contents for CIRRUS photo 600
Page 1: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 User manual...
Page 8: ...000000 2121 813 GA GB 050416...
Page 73: ......
Page 76: ......
Page 77: ...Remote maintenance tool Addendum to the documentation set...
Page 78: ...000000 2121 813 AddGA01 GB 210915...
Page 85: ......
Page 88: ......
Page 104: ...WINDOWS EMBEDDED STANDARD 7 GB 21 06 2012 16...
Page 105: ......