Maintenance and care
000000-2121-813-GA-GB-050416
47
1 Measurement point at head rest (screw)
2 Measurement point at instrument base (metallic connector panel)
3 Measurement point at power supply unit (screw)
Fig. 17
Testing electrical safety
•
After successful measurement, the device leakage current must be
measured. Preferably, the differential current method should be used. The
device is in its operating state. Press the measuring tip onto the
measurement points again (1, 2, 3, Fig. 17). The measured value may not
exceed 0.5 mA.
•
The inspection will be concluded by a function check. This should be
performed by a person who is familiar with the application.
•
Note down the measured values.