Notes on and conditions of use for the remote maintenance tool
000000-2121-813-AddGA-GB-210915
5
Notes on and conditions of use for the remote
maintenance tool
1. Use of the remote maintenance tool requires that the
Carl Zeiss Meditec AG service team or a company authorized by them
be contacted beforehand.
2. The remote maintenance function can be activated by the user only.
This requires the entry of an activation code. The user can then switch
between observation mode and remote control mode. The user can
abort the selected remote maintenance mode at any time by closing the
remote maintenance dialog box. It is not possible for the service techni-
cian to start or restart observation mode without the user's consent.
3. In observation mode, a service technician from Carl Zeiss Meditec AG or
a company authorized by them is able to continuously view a copy of
the current program interface. Service personnel may therefore be able
to view patient details if they are displayed in the visible program inter-
face. It is solely the user’s responsibility to take all necessary measures
to ensure that confidential data is protected during the remote mainte-
nance session and that all legal regulations are adhered to.
4. In remote control mode, a service technician from Carl Zeiss Meditec AG
or an authorized company is able to operate the instrument’s user
interface with the privileges of the currently logged in user. Remote
control mode can be terminated at any time by pressing <F5>. The
program will then switch to observation mode. It is solely the user’s
responsibility to take all necessary measures to ensure that confidential
data is protected during the remote maintenance session and that all
legal regulations are adhered to.
5. The user may not leave the instrument unsupervised in observation or
remote maintenance mode. Under no circumstances is the remote
maintenance tool to be used for fault diagnosis while treatment is being
carried out. The user must alert the service technician from
Carl Zeiss Meditec AG or a company authorized by them by telephone if
there are other people in the room in which the instrument is located
during a remote maintenance session.
6. The user is solely responsible for the safe operation of the instrument
during a remote maintenance session.
Summary of Contents for CIRRUS photo 600
Page 1: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 Documentation set...
Page 4: ......
Page 6: ......
Page 7: ...CIRRUS photo CIRRUS photo 600 and CIRRUS photo 800 User manual...
Page 8: ...000000 2121 813 GA GB 050416...
Page 73: ......
Page 76: ......
Page 77: ...Remote maintenance tool Addendum to the documentation set...
Page 78: ...000000 2121 813 AddGA01 GB 210915...
Page 85: ......
Page 88: ......
Page 104: ...WINDOWS EMBEDDED STANDARD 7 GB 21 06 2012 16...
Page 105: ......