Optional accessories
000000-2121-813-GA-GB-050416
50
Printer
WARNING - RISK OF ELECTRIC SHOCK
The printer must be set up outside the patient environment (1.5 m from the
patient's seat at the instrument), operated with a network isolator and
plugged into a separate socket. The user may not touch the printer and
patient simultaneously.
CAUTION - RISK DUE TO FAULTY DATA
Only a PostScript-capable network printer may be used.
The printer is used for printing capture images. It can be connected to the
CIRRUS photo via the network as a network printer (see Fig. 18) or directly
(peer-to-peer), see the following description.
1 CIRRUS photo
2 Network cable with ferrite core
3 Ferrite core
4 Network isolator
5 Network cable
6 Network printer
Fig. 18
Connection to a printer via a network
The purpose of the network isolator is to provide a galvanic
separation of the network supply (network cable) between the
CIRRUS photo and a network or network printer.
Risks arising from the printer or network for patiants and users are
thereby excluded.