OPERATION
Axioskop 2 FS
plus
Carl Zeiss
Illumination and contrasting techniques
Axioskop 2 FS
MOT
3-22
B 40-076 e 02/01
(4)
Setting the height stop on the
condenser carrier
•
Loosen fixation screw of height stop (3-17/
1
)
using SW 3 ball-headed screwdriver.
•
Use focusing drive to focus on the specimen.
•
Close luminous-field diaphragm and image it in
focus via the height adjustment control
(3-17/
2
) of the universal condenser.
•
Carefully lift the condenser slightly without
touching the specimen.
•
Tighten fixation screw (3-17/
1
) of the height
adjustment.
(5)
Transmitted-light brightfield configuration of the Axioskop 2 FS
MOT
Each Axioskop 2 FS
MOT
configuration permits the performance of the transmitted-light brightfield
technique.
(6)
Setting the Axioskop 2 FS
MOT
for transmitted-light brightfield according to KÖHLER
−
The Axioskop 2 FS
MOT
has been set up as described in chapter 2.
−
Switch on the transmitted-light illumination via the separate power unit.
−
Switch on the control unit for motorized objective focusing.
•
Set KÖHLER illumination as described in section 3.3.1 (3) for the manual Axioskop 2 FS
plus
.
•
Unlike the Axioskop 2 FS
plus
, focusing is made via the keypad of the control unit.
☞
For < 10x objectives, the front lens of the universal condenser must be folded out and the
aperture diaphragm fully opened.
Fig. 3-17
Setting the height stop on the
condenser carrier
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...