START-UP
Axioskop 2 FS
plus
Carl Zeiss
Attachments and conversions
Axioskop 2 FS
MOT
2-28
B 40-076 e 02/01
After expiry of the average operation time of
300 h, the HBO 103 W/2 mercury vapor short-arc
lamp must be exchanged.
The operating time of the lamp can be checked on
the counter (2-37/
1
) of the HBO
100
W
transformer.
Follow the following steps when changing the
HBO 103 W/2 lamp:
•
Remove or unscrew the line plug and
the HBO
103 plug from the HBO
100
W
transformer.
•
Wear protective mask and safety gloves.
•
Use SW 3 ball-headed screwdriver to loosen
clamping screw (2-38/
2
) and remove HBO 103
illuminator (2-38/
3
) from the incident-light tube
(2-38/
1
) of the microscope stand.
•
Use focusing knob (2-39/
3
) to bring collector to
the position at the very front (in the direction of
the light).
•
Use ball-headed screwdriver to loosen clamping
screw for lamp housing (2-39/
2
) and carefully
remove lamp housing (2-39/
1
) from the lamp
mount (2-40/
3
) in an upward direction.
•
Press spring handle (2-40/
4
) down and remove
defective lamp (or transport securing pin) with
the heat sink from the lamp socket 2-40/
2
).
You do not need to remove the cable
connection.
Fig. 2-37
HBO 100 W transformer
Fig. 2-38
Removing the HBO 103 illuminator
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...