START-UP
Axioskop 2 FS
plus
Carl Zeiss
Initial start-up
Axioskop 2 FS
MOT
2-16
B 40-076 e 02/01
2.1.13
HAL 100 halogen illuminator
(1)
Attachment
of
HAL 100
halogen
illuminator
•
Remove cover from transmitted-light tube
(2-17/
3
).
•
Insert dovetail (2-18/
2
) of lamp housing
(2-17/
1
) in transmitted-light tube (2-17/
3
) on
the instrument rear and use the SW 3 ball-
headed screwdriver to tighten clamping screw
(2-17/
2
).
•
Connect 3-pin illuminator plug (2-17/
4
) to
3-pin 12
V/100
W socket (2-14/
1
) of the
separate 12 V DC 100 W power unit (also see
2.1.12).
(2)
Adjustment of HAL
100 halogen
illuminator
a)
Coarse adjustment
•
Loosen clamping screw (2-17/
2
) and remove
the halogen illuminator from the microscope
stand.
•
Switch on separate 12 V DC 100 W power unit
as described under 2.1.12.
•
Direct the light beam against a projection area
(wall) approx. 3 m away.
Do not look into the light exit opening
of the illuminator.
•
Set adjusting screw (2-18/
3
) using SW 3 ball-
headed screwdriver in such a way that both
images of the lamp coil are as sharply focused
as possible on the projection screen.
•
Then set adjusting screws (2-18/
4
and
5
) in
such a way that the lamp coil of one image
exactly fills the gaps in the reflector image
(2-18/
1
).
Fig. 2-17
Attachment of HAL 100 halogen
illuminator
Fig. 2-18
Adjustment of HAL 100 halogen
illuminator
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...