START-UP
Axioskop 2 FS
plus
Carl Zeiss
Initial start-up
Axioskop 2 FS
MOT
2-18
B 40-076 e 02/01
•
Insert multi-pin plug of the HBO 103 illuminator
(2-19/
3
) into socket (2-20/
1
) of the HBO 100 W
transformer and secure it with captive coupling
ring (also see
2.1.12).
•
Connect line cable to socket (2-20/
2
) of the
HBO 100 W transformer first, then connect it to
the line.
(3)
Adjustment of HBO 103
Adjustment of the HBO 103 illuminator described
below is performed in two steps: coarse
adjustment and subsequent fine adjustment. The
latter can be performed with or without adjusting
aid (2-21/
2
) on the incident-light tube (2-21/
1
).
Coarse adjustment:
•
Remove HBO 103 illuminator from the
microscope stand by loosening clamping screw
(2-19/
2
).
•
Switch on HBO 103 on HBO 100 W transformer
(2-16/
1
) and allow it to heat to operating
temperature. Make sure not to look into the
light exit opening of the illuminator.
•
Direct light beam against a projection area
(wall) approx. 3 m away.
•
Use knurled knob (2-22/
6
) to focus on the
brighter focal spot via collector adjustment.
•
Use adjusting screw (2-22/
4
) and SW 3 ball-
headed screwdriver to locate and focus on the
slightly darker, reflected focal spot.
•
Use adjusting screws (2-22/
1
and
3
) and SW 3
ball-headed screwdriver to set the height and
side of the brighter focal spot in the same way
as the focal spot image (2-24).
Fig. 2-20
HBO 100 W transformer
Fig. 2-21
Adjusting aid
Fig. 2-22
Adjustment of mercury vapor short-
arc lamp
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...