![Zeiss Axioskop 2 FS mot Manual Download Page 115](http://html1.mh-extra.com/html/zeiss/axioskop-2-fs-mot/axioskop-2-fs-mot_manual_927664115.webp)
Axioskop 2 FS
plus
CARE, MAINTENANCE, TROUBLESHOOTING ...
Axioskop 2 FS
MOT
Troubleshooting
Carl Zeiss
B 40-076 e 02/01
4-5
4.3
Troubleshooting
Problem
Cause
Remedy
Shadows or inhomogeneous
image brightness in the field of
view; the field is not entirely
visible.
The vis/phot pushrod on the
camera tube is not in the correct
(intermediate) position.
Move the vis/phot pushrod on
the phototube in the correct
position (end position),
see page 3-10.
Objective in objective slider not
lowered to work position.
Lower objective in objective slider
to work position.
Condenser not set correctly.
Set condenser correctly
(adjustment, centering),
see p. 3-20 ff.
Aperture diaphragm not set
correctly.
Set aperture diaphragm correctly
(centering, opening),
see p. 3-20 ff.
Luminous-field diaphragm not set
correctly.
Set luminous-field diaphragm
correctly (centering, opening),
see p. 3-20 ff.
Filter not inserted correctly in filter
mount.
Insert filter correctly in filter
mount.
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...