![Zeiss Axioskop 2 FS mot Manual Download Page 118](http://html1.mh-extra.com/html/zeiss/axioskop-2-fs-mot/axioskop-2-fs-mot_manual_927664118.webp)
CARE, MAINTENANCE, TROUBLESHOOTING ...
Axioskop 2 FS
plus
Carl Zeiss
Troubleshooting
Axioskop 2 FS
MOT
4-8
B 40-076 e 02/01
Problem
Cause
Remedy
The 12 V 100 W halogen lamp
does not function although the
on / off switch is in "on"
position.
Line cable not connected to the
line.
Connect line cable to the line and
make sure to check the instrument
and line voltage.
12 V 100 W halogen lamp not
installed.
Attach 12 V 100 W halogen lamp,
see p. 2-27.
12 V 100 W halogen lamp
defective.
Replace 12 V 100 W halogen
lamp, see p. 2-27.
The specified 12 V 100 W halogen
lamp is not used.
Use the specified 6 V 12 W
halogen lamp, see p. 1-6.
Fuses are defective.
Exchange fuses, see page 4-4.
The 12 V 100 W halogen lamp
flickers, unstable brightness.
End of average life of 12 V 100 W
halogen lamp.
Replace 12 V 100 W halogen
lamp, see p. 2-27.
Incorrectly installed or broken line
cable.
Connect line cable correctly or
replace it.
The pins of the 12 V 100 W
halogen lamp are not correctly
inserted into the receptacle.
Correctly insert pins of 12 V
100 W halogen lamp in
receptacle, see p. 2-27.
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...