Axioskop 2 FS
plus
START-UP
Axioskop 2 FS
MOT
Initial start-up
Carl Zeiss
B 40-076 e 02/01
2-17
b)
Fine adjustment
•
Attach illuminator to the stand again and tighten clamping screw (2-17/
2
).
•
Remove diffusion disk (3-1/
12
) and filters (3-1/
15
), if switched in, from the beam path.
•
Focus on the specimen using objective
≤
40x and search for a free object area.
•
Remove eyepiece and center lamp coil and its reflector image in the pupil image using adjusting
screws (2-18/
4
and
5
).
•
Use adjusting screws (2-18/
3
) to optimize the homogeneous illumination of the pupil image.
•
Switch in diffusion disk and required filters.
2.1.14
HBO 103 illuminator
(1)
Attach HBO 103 W/2 mercury vapor
short-arc lamp
The HBO 103 illuminator and the HBO 103 W/2
mercury vapor short-arc lamp are supplied in
separate packages for safety reasons.
Therefore, insertion of the HBO 103 W/2 lamp into
the lamp housing is the first step to the start-up of
this illuminator.
Insertion of a new HBO 103 W/2 lamp is described
in detail on page 2-27, section 2.2.8 entitled
"Changing the HBO 103 W/2 mercury vapor short-
arc lamp".
(2)
Attachment of HBO 103 illuminator
•
Remove cover from the incident-light tube
(2-19/
1
).
•
Insert dovetail of lamp housing in incident-light
tube (2-19/
1
) on the instrument rear and use
SW
3 ball-headed screwdriver to tighten
clamping screw (2-19/
2
).
Fig. 2-19
Attachment of HBO 103 illuminator
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...