Axioskop 2 FS
plus
START-UP
Axioskop 2 FS
MOT
Attachments and conversions
Carl Zeiss
B 40-076 e 02/01
2-29
•
Loosen the fixing screw (2-40/
5
) on the heat
sink (2-40/
6
)
and remove heat sink. If you
cannot remove (or replace) the heat sink, take
out the screw completely, screw it into the
neighboring threaded hole and use it to press
out the heat sink.
•
Dispose of defective lamp according to
regulations.
•
Place new lamp with the smaller diameter into
the heat sink until it touches.
When inserting, make sure to use
the hole with the correct diameter
and do not put any pressure on the
lamp.
Correct placement in the heat sink is
indicated with
H
(for HBO) or
X
(for
XBO).
•
Clamp fixing screw onto heat sink.
•
Place new lamp (2-40/
1
) with heat sink into the
appropriate position in the lamp socket (2-40/
2
)
while pressing down on the spring handle
(2-40/
4
). Make sure that the feed line to the
heat sink is free on the side. Release spring
handle. Do not put any pressure on the lamp.
•
Place lamp housing (2-39/
1
) on the lamp mount
(2-40/
3
) and tighten with clamping screw
(2-39/
2
).
•
Connect HBO 103 plug and line plug to the
HBO 100 W transformer.
•
Adjust HBO 103 illuminator as described in
section 2.1.14 (3).
Fig. 2-39
Removing HBO 103 lamp housing
Fig. 2-40
Changing the HBO 103 W/2 mercury
pressure short-arc lamp
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...