START-UP
Axioskop 2 FS
plus
Carl Zeiss
Attachments and conversions
Axioskop 2 FS
MOT
2-38
B 40-076 e 02/01
•
Now tip up the module half on top (2-52/
1
)
and remove from the retaining elements
(2-52/
5b
) of the bottom module half.
•
Remove the beam splitter (2-52/
2
) and spring
frame (2-52/
3
) from the bottom module half.
•
Remove the old beam splitter and carefully
place the new one with the reflective side
facing downwards on the spring frame (2-52/
4
)
and place both parts together in the bottom
module half. Make sure that the side latch of
the spring frame is in the appropriate recess in
the bottom module half.
The reflective side of the beam splitter is
recognizable by carefully placing a wooden splint
on the surface of the beam splitter. There is no
spacing between the splint and its mirror image
•
Place the top module half (2-52/
1
) on the
bottom module half (2-52/
4
) (the retaining
elements 2-52/
5b
grip the eyes 2-52/
5a
). Hold
both halves together and turn back to the
installation position
•
Re-insert the slotted screws and tighten them.
•
Finally, attach the adhesive label with the name
of the filter combination to the side of the
module.
Fig. 2-52
Changing the beam splitter
Summary of Contents for Axioskop 2 FS mot
Page 1: ...Operating Manual Axioskop 2 FS plus FS MOT Research Microscope...
Page 14: ...INTRODUCTION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 0 14 B 40 076 e 02 01...
Page 110: ...OPERATION Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT 3 40 B 40 076 e 02 01...
Page 122: ...ANNEX Axioskop 2 FS plus Carl Zeiss Axioskop 2 FS MOT A 2 B 40 076 e 02 01...