INTRODUCTION
Axio Zoom.V16
Notes on instrument safety
Carl Zeiss
05/2012
435080-9030-701
9
On microscopes with motorized focusing drive there is a
risk of pinching fingers
in the
working area when lowering the body of the microscope.
−
Do not reach into the work area or beneath the motorized focusing drive during lowering.
−
Automatic travel can be interrupted by pressing the STOP button on the focusing drive
(Fig. 28/
1
, page 43) or by moving the knurled knob on the HIP up and down.
−
The lower limit switch of the motorized focusing drive must always be set to prevent the
objective from colliding with the stage equipment/specimen (see Section 4.6, page 44).
The instruments are not equipped with any special devices for protection against substances
that are corrosive, potentially infectious, toxic or radioactive, or other substances that could be
hazardous to health. All legal regulations must be observed when handling such substances,
particularly the relevant national accident prevention regulations.
Do not operate the supplied equipment
- in areas in which there is a danger of explosion,
- in the presence of volatile anaesthetics or flammable solvents, such as alcohol, benzene, etc.
Do not switch on the instrument unless all cables have been connected; switch it off before
disconnecting any cables.
If more than one CAN component is used, these can be connected either to each other or to
the focus motor. In such cases only
one
power supply unit may be connected to the
microscope system.
VisLED annular lights S are LED Class 1 equipment. Do not look directly into the LED light. To
avoid glare during transmitted light applications, swivel in anti-glare shield (Fig. 3/
I
, page 12).
If the device is operated with an external cold light source (high-energy light) never look directly
into the light guide exit of the cold light source. There is otherwise a risk of dazzling or
blinding.
The lamps on the light guide source must be replaced in accordance with the manufacturer's
instructions. If the instructions are not complied with there is a risk of burning or explosion
when the lamps are replaced.