SET-UP
Carl Zeiss
Mounting the stages
Axio Zoom.V16
26
435080-9030-701
05/2012
3.7
Mounting the stages
Remove the shipping braces from the stages before assembly.
3.7.1
Adapter ring for stand base 450
When using stand base 450 it must be
ensured that the adapter ring
(Fig. 12/
3
) has been mounted for
holding stages.
•
Removing stage insert plate (Fig. 12/
2
). Remove
the three screws (Fig. 12/
1
).
•
Insert the adapter ring (Fig. 12/
3
) into the stand
base(Fig. 12/
4
) and secure with the three
captive screws (Fig. 12/
5
).
•
Attach the stage insert plate and screw it tight.
This adapter ring is already integrated
into stand base 300 (base 300
for
reflected light or transillumination
base 300).
3.7.2
Adapter for stages with 84 mm and
120 mm mounting faces
Different adapters are required for mounting
stages with an 84 mm mounting face:
Stand base
Stage with mounting face
84 mm
120 mm
Stand base 450
Adapter
84/120
No adapter
Base 300 for
reflected light
Adapter
84/120
No adapter
Transillumination
base 300
Adapter
84+120
Adapter
84+120
Stand base 450
•
Remove the three short socket head screws
(SW 3) from the insert plate (Fig. 13/
3
) of stand
base 450 (Fig. 13/
4
).
•
Screw on adapter 84/120 (Fig. 13/
5
) with the
three longer socket head screws through the
insert plate and hand-tighten.
Base 300
for reflected light
and transillumination base 300
•
Screw adapter 84/120 (Fig. 13/
5
) directly to the base 300
for
reflected light or adapter 84+120
5
) to transillumination base 300 (Fig. 13/
2
) and hand-tighten.
Fig. 12 Mounting the adapter ring
Fig. 13 Mounting adapter 84/120 or 84+120
Summary of Contents for Axio Zoom.V16
Page 1: ...Operating Manual Axio Zoom V16 Zoom microscope...
Page 2: ......