Ambient conditions
Storage (in packaging)
Permissible ambient temperature ....................................................................................... +10 to +40 °C
Permissible relative humidity ....................................................... max. 75 % at +35 °C (non-condensing)
Transport (in packaging):
Permissible ambient temperature ......................................................................................... -40 to +70°C
Operation
Permissible ambient temperature ....................................................................................... +10 to +40 °C
Permissible relative humidity .................................................................................................. max. 75 %
Air pressure ............................................................................................................. 800 hPa to 1060 hPa
Degree of pollution ............................................................................................................................... 2
Operational area ................................................................................................................ Closed rooms
Highest permitted altitude of use ......................................................................................... max. 2000 m
Operating data
Plug-in power supply as a component of HIP
Electrical protection class ...................................................................................................................... II
Ingress protection rating .................................................................................................................. IP 40
Mains voltage (wide range) ..................................................................................... 100 to 240 V ±10 %
Mains frequency .................................................................................................................... 50 to 60 Hz
Power consumption ................................................................................................................... 700 mA
Output voltage ..................................................................................... stabilized 24 V DC, 1.25 A, 30 W
RFI suppression...................................................................................... conforming to EN 55011 Class A
Noise immunity ........................................................................................ conforming to DIN EN 61326-1
Electrical safety..................................................................................... to DIN EN 61010-1 (IEC 61010-1)
................................................................................................
conforming to CSA and UL requirements
Tabletop power supply unit for transmitted light illumination
Mains voltage (wide range) ..................................................................................... 100 to 240 V ±10 %
Mains frequency .................................................................................................................... 50 to 60 Hz
Power consumption ....................................................................................................................... 1.4 A