SET-UP
Carl Zeiss
Inserting the VisiLED transmitted light unit H
Axio Zoom.V16
34
435080-9030-701
05/2012
Base 300 for reflected light
•
If necessary, remove the glass plate (Fig. 22/
1
) from the base 300
for reflected light
.
•
Insert the VisiLED transmitted light unit H (Fig. 22/
3
) into the base 300
for reflected light
2
) from
below.
•
Insert the cable into the base 300
f
or transmitted light and pull the plug through to the outside
A
).
•
Secure the VisiLED transmitted light unit H with a clamp screw (Fig. 22/
4
).
•
Insert the glass plate (Fig. 22/
1
) into the base 300
for reflected light
.
•
Connect the VisiLED transmitted light unit to the controller MC 1000 and plug into the respective
power supply unit (see separate operating manual).
Fig. 22 Inserting the VisiLED transmitted light unit H, d=84 mm into the base 300 for reflected light