![background image](http://html2.mh-extra.com/html/zeiss/axio-zoom-v16/axio-zoom-v16_operating-manual_3350140025.webp)
SET-UP
Axio Zoom.V16
Mounting and connecting Manual Rotary Control (MaRC)
Carl Zeiss
05/2012
435080-9030-701
25
•
Attach the mounting plate (Fig. 9/
2
) with a screw (Fig. 9/
6
) securely to the flange of the focusing drive
1
) (or microscope body). The screw must be inserted into the left hole of the mounting plate.
Please ensure that the orientation pin (Fig. 9/
3
) of the mounting plate is seated in the groove provided
9
).
•
Screw the MaRC to the attached mounting plate (Fig. 9/
2
) with a socket head screw (SW 3; Fig. 9/
4
)
using a ball-head screwdriver (Fig. 9/
5
).
•
Plug the
CAN cable RJ 45
into one of the two CAN bus ports on the MaRC (base) and connect it to the
motorized focusing drive.
Mounting the MaRC on the table-top base:
•
Secure the mounting plate (Fig. 10/
2
) with
screw (Fig. 10/
3
) on the table-top base
1
). The screw must be inserted into the
right hole of the mounting plate.
•
Screw the MaRC (Fig. 10/
4
) to the attached
mounting plate with a socket head screw
(SW 3; Fig. 10/
5
) using a ball-head screwdriver
6
).
•
Plug the
CAN cable RJ 45
3
) into one of
the two CAN bus ports of the MaRC (through
the underside of the table-top base, Fig. 11/
1
)
and connect to the motorized focusing drive.
•
Pull the
CAN cable RJ 45
to the outside through
one of the two cut-outs (right or left, Fig. 11/
2
).
•
Place the table-top base with MaRC beside the
microscope to the right or left.
Fig. 10 Mounting the MaRC on the table-top
base
Fig. 11 Connecting CAN cable RJ 45 to MaRC
with table-top base