SET-UP
Axio Zoom.V16
Mounting the microscope
Carl Zeiss
05/2012
435080-9030-701
21
•
Screw the objective (Fig. 6/
10
) counterclockwise into the body of the microscope until it engages.
Screw a second objective (Fig. 6/
11
) into the nosepiece as required.
•
Insert the tubes (Fig. 6/
1
,
2
,
3
) or intermediate tubes (Fig. 6/
5
) with dovetail mount into the body of
the microscope and hand-tighten it with a socket head screw (Fig. 6/
7
or
4
) using a ball-head
screwdriver.
If the binocular ergo phototube Z is used, remove the shipping screw (socket head screw SW 3)
on the underside and store it for use at a later date.
Either an intermediate tube (spacer tube, Fig. 6/
5
) or Fluar Illuminator Z mot.) may be mounted
between the body of the microscope and the binocular phototube. This increases the eyepoint
height by 40 mm.
Should the microscope mounted on the manual focusing drive slip under its own weight, please
adjust the torque (see page 19).