Version 7.0
G-30-1781-en
Page 183
S100 / OPMI pico
Operation
A target value in a range from 50% to 70% usually provides the best ex-
posure results.
Low light
This mode has the same function as the "Automatic" mode, but exposure
times greater than 1/50 (60) s are allowed, which can result in blurred
movements.
Manual
In this mode you can set the exposure time under "Exposure time". The
setting range extends from 1/6 to 1/10000.
– Increments for PAL: 1/10000, 1/4000, 1/2000, 1/1000, 1/500,
1/250, 1/100, 1/50, 1/25, 1/12 and 1/6
– Increments for NTSC: 1/10000, 1/4000, 1/2000, 1/1000, 1/500,
1/250, 1/120, 1/60, 1/30, 1/15 and 1/8
2
Lightmeter
In "Lightmeter" you can choose between three types of exposure meas-
urement: Small spot, large spot or integral.
The lightmeter function is active only when the brightness is to set "Auto-
matic" or "Little light".
Small spot
The exposure is measured in a very small area in the image center. This
measurement is suitable for work in narrow channels when an extremely
small illuminated-field diameter is being used.
Large spot
The exposure is measured in an area in the image center. This measure-
ment is ideal for working with an eclipsed surgical field edge (resulting
from a reduced illuminated-field diameter).
Integral
The exposure is measured and averaged across the full video image. This
measurement is recommended for surgical fields that are fully and evenly
lit.
If the object of interest to you is not in the image center, in most cases the
"Small spot" setting will not deliver the desired exposure result. Select ei-
ther a larger area ("Integral" or "Large spot") or manual exposure.
Fig. 80: Types of
exposure measurement
Integral
Large spot
Small spot
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...