What to do in the event of malfunctions
S100 / OPMI pico
Version 7.0
Page 220
G-30-1781-en
Fault in the LED lilght source
The surgical microscope is equipped with an LED light source.
1
LED error message
If the amber LED is illuminated, the LED light source is in a limited oper-
ating mode.
In this case, the light intensity drops down to 50 %.
• Please contact ZEISS Service.
Fig. 100: Malfunction in the
LED light source
Failure concept for the LED light source
Automatic treatment continuation due to integrated failure concept:
– The LED light source uses an integrated failure concept for automatic
treatment continuation without an interruption.
– Individual LEDS are electronically arranged in two independent strands,
where one strand automatically continues to be use in the event of a mal-
function.
50% light intensity
1
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...