What to do in the event of malfunctions
S100 / OPMI pico
Version 7.0
Page 214
G-30-1781-en
No image, video indicator
lamp is lit
If a video recorder is con-
nected between CCU and
monitor: Defective video re-
corder.
Connect CCU and monitor di-
rectly.
Defective CCU or camera
head in surgical microscope,
or defective connecting cable
between CCU and surgical
microscope.
Contact service.
–
Image too dark,
snow effect on monitor.
Monitor not correctly set.
Set monitor to default status, or
increase brightness and contrast.
Please refer
to monitor's
Instructions
for Use.
Brightness of video unit set
too low.
Increase brightness of video unit. Page 167
Surgical field illumination too
dark.
Increase brightness of illumina-
tion.
Image too bright,
Milky image on monitor
Monitor not correctly set.
Set monitor to default status, or
reduce brightness and/or con-
trast.
Please refer
to monitor's
Instructions
for Use.
Brightness of video unit is set
too high.
Reduce brightness of video unit. Page 167
Surgical field illumination too
bright.
Reduce brightness of surgical
field illumination.
ENT application: Image de-
tail in black speculum too
bright, too shiny
Video unit: Lightmeter function - set "Small Spot"
Page 82
Brightness of CCU set too
high.
Reduce brightness of video unit. Page 167
Brightness control is based on
black color of speculum.
Center the speculum and object
in center of screen.
–
Use a metal speculum (gray,
stainless steel).
–
Problem
Possible cause
Remedy
Cross
reference
Summary of Contents for 000000-1080-349
Page 1: ...ZEISS S100 OPMI pico Instructions for Use G 30 1781 en Version 7 0 2015 09 09...
Page 4: ...S100 OPMI pico Version 7 0 Page 4 G 30 1781 en...
Page 6: ...S100 OPMIpico Version 7 0 Page 6 G 30 1781 en...
Page 49: ...Version 7 0 G 30 1781 en Page 49 S100 OPMI pico Safety Measures Fig 21 Video units 1 2 4 1 2 3...
Page 50: ...Safety Measures S100 OPMI pico Version 7 0 Page 50 G 30 1781 en...
Page 88: ...System Overview S100 OPMI pico Version 7 0 Page 88 G 30 1781 en...
Page 115: ...Version 7 0 G 30 1781 en Page 115 S100 OPMI pico Preparations for use...
Page 126: ...Preparations for use S100 OPMI pico Version 7 0 Page 126 G 30 1781 en...
Page 169: ...Version 7 0 G 30 1781 en Page 169 S100 OPMI pico Operation...
Page 175: ...Version 7 0 G 30 1781 en Page 175 S100 OPMI pico Operation...
Page 264: ...Indexes S100 OPMI pico Version 7 0 Page 264 G 30 1781 en...
Page 276: ...S100 OPMI pico Version 7 0 Page 276 G 30 1781 en...
Page 277: ...Version 7 0 G 30 1781 en Page 277 S100 OPMI pico Blank page for your notes...