5-76
IM DLM4038-17EN
:ANALysis:PMEAsure<x>:SETup:I?
Function
Queries all current input channel settings for
power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:I?
<x> = 1 to 4
Example
:ANALYSIS:PMEASURE1:SETUP:I?
:ANALysis:PMEAsure<x>:SETup:I:DESKew
Function
Sets or queries the auto deskewing of the current
input channel for power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:I:DESK
ew {<Time>}
:ANALysis:PMEAsure<x>:SETup:I:DESK
ew?
<x> = 1 to 4
<Time> = −100.0 ns to 100.0 ns (in 10-ps steps)
Example
:ANALYSIS:PMEASURE1:SETUP:I:DESK
EW 1NS
:ANALYSIS:PMEASURE1:SETUP:I:DESKEW?
-> :ANALYSIS:PMEASURE1:SETUP:I:DESK
EW 1.000E−09
:ANALysis:PMEAsure<x>:SETup:I:PROBe
Function
Queries all current-to-voltage conversion ratio
settings of the probe connected to the current
input channel for power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:I:
PROBe {C0_001|C0_002|C0_005|C0_01|
C0_02|C0_05|C0_1|C0_2|C0_5|C1|C2|C5|
C10|C20|C50|C100|C200|C500|C1000|
C2000}
:ANALysis:PMEAsure<x>:SETup:I:PROBe?
<x> = 1 to 4
Example
:ANALYSIS:PMEASURE1:SETUP:I:PROBE C0
_001
:ANALYSIS:PMEASURE1:SETUP:I:PROBE?
-> :ANALYSIS:PMEASURE1:SETUP:I:PROB
E C0_001
:ANALysis:PMEAsure<x>:SETup:RTRace
Function
Sets or queries the deskewing source trace for
power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:RTRa
ce {I|U}
:ANALysis:PMEAsure<x>:SETup:RTRace?
<x> = 1 to 4
Example
:ANALYSIS:PMEASURE1:SETUP:RTRACE I
:ANALYSIS:PMEASURE1:SETUP:RTRACE?
-> :ANALYSIS:PMEASURE1:SETUP:RTRA
CE I
:ANALysis:PMEAsure<x>:SETup:U?
Function
Queries all voltage input channel settings for
power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:U?
<x> = 1 to 4
Example
:ANALYSIS:PMEASURE1:SETUP:U?
:ANALysis:PMEAsure<x>:SETup:U:DESKew
Function
Sets or queries the auto deskewing of the voltage
input channel for power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:U:DESK
ew {<Time>}
:ANALysis:PMEAsure<x>:SETup:U:DESK
ew?
<x> = 1 to 4
<Time> = −100.0 ns to 100.0 ns (in 10-ps steps)
Example
:ANALYSIS:PMEASURE1:SETUP:U:DESK
EW 1NS
:ANALYSIS:PMEASURE1:SETUP:U:DESKEW?
-> :ANALYSIS:PMEASURE1:SETUP:U:DESK
EW 1.000E−09
:ANALysis:PMEAsure<x>:SETup:U:PROBe
Function
Queries all voltage-to-voltage conversion ratio
settings of the probe connected to the voltage
input channel for power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:U:PRO
Be {<NRf>}
:ANALysis:PMEAsure<x>:SETup:U:PROBe?
<x> = 1 to 4
<NRf> = See the DLM4000 Features Guide for
this information.
Example
:ANALYSIS:PMEASURE1:SETUP:U:PROBE 1
:ANALYSIS:PMEASURE1:SETUP:U:PROBE?
−> :ANALYSIS:PMEASURE1:SETUP:U:PROB
E 1.000
:ANALysis:PMEAsure<x>:TRANge
Function
Sets or queries the measurement time period.
Syntax
:ANALysis:PMEAsure<x>:TRANge {<NRf>,
<NRf>}:
ANALysis:PMEAsure<x>:TRANge?
<x> = 1 to 4
<NRf>, <NRf> = −5 to 5 divisions (in steps of 10
divisions/display record length)
Example
:ANALYSIS:PMEASURE1:TRANGE 5,−5
:ANALYSIS:PMEASURE1:TRANGE?
-> :ANALYSIS:PMEASURE1:TRANGE 5.0000
000E+00,−5.000
:ANALysis:PMEAsure<x>:UDPRoximal?
Function
Queries all distal, mesial, and proximal settings.
Syntax
:ANALysis:PMEAsure<x>:UDPRoximal?
<x> = 1 to 4
Example
:ANALYSIS:PMEASURE1:UDPROXIMAL?
5.3 ANALysis Group