5-145
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure?
Function
Queries all of the settings for measuring the delay
between channels of a logic waveform in the
specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure?
<x> = 0 to 7
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure:COU
Nt
Function
Sets or queries the count number of a source
waveform for measuring the delay between
channels of a logic waveform in the specified
area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure:
COUNt {<NRf>}
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure:
COUNt?
<x> = 0 to 7
<NRf> = 1 to 10
Example
:MEASURE:PODL1:AREA1:DELAY:MEASURE:C
OUNT 2
:MEASURE:PODL1:AREA1:DELAY:MEASURE:C
OUNT?
-> :MEASURE:PODL1:AREA1:DELAY:MEASUR
E:COUNT 2
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure:SLO
Pe
Function
Sets or queries the slope of a source waveform
for measuring the delay between channels of a
logic waveform in the specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure:SLO
Pe {FALL|RISE}
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:MEASure:SLO
Pe?
<x> = 0 to 7
Example
:MEASURE:PODL1:AREA1:DELAY:MEASURE:S
LOPE FALL
:MEASURE:PODL1:AREA1:DELAY:MEASURE:S
LOPE?
-> :MEASURE:PODL1:AREA1:DELAY:MEASUR
E:SLOPE FALL
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence?
Function
Queries all reference waveform settings used to
measure the delay between the specified logic
waveform and the reference waveform in the
specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence?
<x> = 0 to 7
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:CO
UNt
Function
Sets or queries the count number of a reference
waveform for measuring the delay between
channels of a logic waveform in the specified
area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:
COUNt {<NRf>}
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:
COUNt?
<x> = 0 to 7
<NRf> = 1 to 10
Example
:MEASURE:PODL1:AREA1:DELAY:
REFERENCE:COUNT 2
:MEASURE:PODL1:AREA1:DELAY:
REFERENCE:COUNT?
-> :MEASURE:PODL1:AREA1:DELAY:REFERE
NCE:COUNT 2
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:SL
OPe
Function
Sets or queries the slope of the reference
waveform used to measure the delay between
the specified logic waveform and the reference
waveform in the specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:
SLOPe {FALL|RISE}
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:
SLOPe?
<x> = 0 to 7
Example
:MEASURE:PODL1:AREA1:DELAY:
REFERENCE:SLOPE FALL
:MEASURE:PODL1:AREA1:DELAY:
REFERENCE:SLOPE?
-> :MEASURE:PODL1:AREA1:DELAY:REFERE
NCE:SLOPE FALL
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:SO
URce
Function
Sets or queries whether the reference point for
measuring the delay between channels of a logic
waveform will be set to the trigger point or a
waveform in the specified area.
Syntax
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:
SOURce {TRACe|TRIGger}
:MEASure:{PODA<x>|PODB<x>|PODL<x>}
{[:AREA1]|:AREA2}:DELay:REFerence:
SOURce?
<x> = 0 to 7
Example
:MEASURE:PODL1:AREA1:DELAY:
REFERENCE:SOURCE TRACE
:MEASURE:PODL1:AREA1:DELAY:
REFERENCE:SOURCE?
-> :MEASURE:PODL1:AREA1:DELAY:REFERE
NCE:SOURCE TRACE
5.21 MEASure Group