5-209
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CRCType
Function
Sets or queries the SENT signal analysis CRC
computation type.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CRCType {LEGacy|RECommended}
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CRCType?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CRCTYPE LEGACY
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CRCTYPE?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FO
RMAT:CRCTYPE LEGACY
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CTICk
Function
Sets or queries the SENT signal analysis clock
tick value.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CTICk {<NRf>}
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CTICk?
<x> = 1 to 4
<NRf> = 1.00 µs to 100.00 µs (resolution:0.01 µs)
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CTICK 0.000001
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CTICK?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FO
RMAT:CTICK 1.000000E-06
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CTOLerance
Function
Sets or queries the SENT signal analysis clock
tolerance.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CTOLerance {<NRf>}
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:CTOLerance?
<x> = 1 to 4
<NRf> = ±1.0% to ±30.0% (resolution:0.1%)
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CTOLerance 25.0
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:CTOLERANCE?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FO
RMAT:CTOLERANCE 25.0E+00
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:DNIBbles
Function
Sets or queries the number of data nibbles for
SENT signal analysis.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:DNIBbles {<NRf>}
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:DNIBbles?
<x> = 1 to 4
<NRf> = 1 to 6
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:DNIBBLES 6
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:DNIBBLES?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FO
RMAT:DNIBBLES 6
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:PPULse
Function
Sets or queries the presence or absence of
pause pulses for SENT signal analysis.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:PPULse {<Boolean>}
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:PPULse?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:PPULSE ON
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:PPULSE?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FO
RMAT:PPULSE 1
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:VERSion
Function
Sets or queries the SENT signal analysis
specification version.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:VERSion {APR2016|FEB2008|
JAN2010}
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:VERESion?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:VERSION JAN2010
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMA
T:VERSION?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FO
RMAT:VERSION JAN2010
5.25 SERialbus Group