5-302
IM DLM4038-17EN
:TRIGger[:ATRigger]:OR:ALL
Function
Sets or queries whether to make all channels
edge OR trigger sources at once.
Syntax
:TRIGger[:ATRigger]:OR:
ALL {<Boolean>}
:TRIGger[:ATRigger]:OR:ALL?
Example
:TRIGGER:ATRIGGER:OR:ALL ON
:TRIGGER:ATRIGGER:OR:ALL?
-> :TRIGGER:ATRIGGER:OR:ALL 1
:TRIGger[:ATRigger]:OR:CHANnel<x>
Function
Sets or queries the slope of each channel of the
edge OR trigger.
Syntax
:TRIGger[:ATRigger]:OR:CHANnel
<x> {OFF|RISE|FALL|IN|OUT|BOTH}
:TRIGger[:ATRigger]:OR:CHANnel<x>?
<x> = 1 to 8
Example
:TRIGGER:ATRIGGER:OR:CHANNEL1 RISE
:TRIGGER:ATRIGGER:OR:CHANNEL1?
-> :TRIGGER:ATRIGGER:OR:CHANNEL1 RIS
E
Description If
:TRIGger:SOURce:CHANnel<x>:WINDo
w
is ON, select from
{OFF|IN|OUT}
; otherwise
select from
{OFF|RISE|FALL|BOTH}
.
:TRIGger[:ATRigger]:PSI5?
Function
Queries all PSI5 signal trigger settings.
Syntax
:TRIGger[:ATRigger]:PSI5?
:TRIGger[:ATRigger]:PSI5:DATA?
Function
Queries all PSI5 signal trigger data settings.
Syntax
:TRIGger[:ATRigger]:PSI5:DATA?
:TRIGger[:ATRigger]:PSI5:DATA:CONDit
ion
Function
Sets or queries the data condition for PSI5 signal
triggering.
Syntax
:TRIGger[:ATRigger]:PSI5:DATA:CONDit
ion {EQUal|TRUE}
:TRIGger[:ATRigger]:PSI5:DATA:CONDit
ion?
Example
:TRIGGER:ATRIGGER:PSI5:DATA:CONDITI
ON EQUAL
:TRIGGER:ATRIGGER:PSI5:DATA:CONDITI
ON?
-> :TRIGGER:ATRIGGER:PSI5:DATA:CONDI
TION EQUAL
:TRIGger[:ATRigger]:PSI5:DATA:DECim
al
Function
Sets or queries the data for PSI5 signal triggering
in decimal notation.
Syntax
:TRIGger[:ATRigger]:PSI5:DATA:DECim
al {<NRf>}
:TRIGger[:ATRigger]:PSI5:DATA:DECim
al?
When <NRf> = 10 bit data, −512 to 511
16 bit data, −32768 to 32767
Example
:TRIGGER:ATRIGGER:PSI5:DATA:DECIM
AL −10
:TRIGGER:ATRIGGER:PSI5:DATA:DECIMAL?
-> :TRIGGER:ATRIGGER:PSI5:DATA:DECIM
AL −10
Description This command is valid when
:TRIGger[:ATRig
ger]:PSI5:DATA:CONDition
is
EQUal
.
:TRIGger[:ATRigger]:PSI5:DATA:HEXa
Function
Sets the data for PSI5 signal triggering in
hexadecimal notation.
Syntax
:TRIGger[:ATRigger]:PSI5:DATA:HE
Xa {<String>}
Example
:TRIGGER:ATRIGGER:PSI5:DATA:HE
XA "3A0F"
Description This command is valid when
:TRIGger[:ATRig
ger]:PSI5:DATA:CONDition
is
TRUE
.
:TRIGger[:ATRigger]:PSI5:DATA:PATTe
rn
Function
Sets or queries the data for PSI5 signal triggering
in binary notation.
Syntax
:TRIGger[:ATRigger]:PSI5:DATA:PATTe
rn {<String>}
:TRIGger[:ATRigger]:PSI5:DATA:PATTe
rn?
Example
:TRIGGER:ATRIGGER:PSI5:DATA:PATTE
RN "0011010100001111"
:TRIGGER:ATRIGGER:PSI5:DATA:PATTERN?
-> :TRIGGER:ATRIGGER:PSI5:DATA:PATTE
RN "0011010100001111"
Description This command is valid when
:TRIGger[:ATRig
ger]:PSI5:DATA:CONDition
is
TRUE
.
:TRIGger[:ATRigger]:PSI5:DATA:PFORm
at
Function
Sets or queries the data input format for PSI5
signal triggering.
Syntax
:TRIGger[:ATRigger]:PSI5:DATA:PFORm
at {BINary|HEXa}
:TRIGger[:ATRigger]:PSI5:DATA:PFORm
at?
Example
:TRIGGER:ATRIGGER:PSI5:DATA:PFORM
AT BINARY
:TRIGGER:ATRIGGER:PSI5:DATA:PFORMAT?
-> :TRIGGER:ATRIGGER:PSI5:DATA:PFORM
AT BINARY
5.34 TRIGger Group